Raman Peak Identification Table — Common Raman Shifts & Band Assignments

This searchable Raman peak identification table covers over 80 common Raman spectroscopy peaks organized by material class — organic functional groups, aromatic compounds, polymers, carbon materials, minerals, biological molecules, solvents, and gases. Search by assignment name, material, or wavenumber to instantly filter results. Click any row to see detailed notes and a visual position bar.

The same data powers SpectralBench's Raman Peak Identifier — the automated tool that detects peaks in your uploaded Raman spectrum and matches them against known assignments.

Common Raman Shifts — Quick Reference Table
Material / BondRaman Shift (cm⁻¹)Relative IntensityAssignment
Anatase (TiO₂)144StrongEg mode — distinguishes anatase from rutile
Quartz (SiO₂)464StrongSi-O-Si bending — common mineral marker
S-S disulfide500–550StrongDisulfide stretch — proteins, rubber (IR-inactive)
Crystalline silicon520StrongFirst-order optical phonon — calibration standard
Tryptophan759StrongIndole ring breathing — protein marker
Polypropylene809StrongCH₃ rocking — polymer identification
Hydroxyapatite (bone)960StrongSymmetric PO₄³⁻ stretch — bone mineral marker
Sulfate (SO₄²⁻)983StrongSymmetric stretch — free sulfate ion
Polystyrene / benzene ring1001StrongRing breathing — calibration standard
Calcite (CaCO₃)1086StrongSymmetric CO₃²⁻ stretch — carbonate minerals
D band (carbon)1350VariableDisorder/defect band — carbon crystallinity
Carotenoid1525StrongC=C stretch — resonance-enhanced biological pigment
G band (graphitic carbon)1580StrongE₂g mode of sp² carbon — graphene, nanotubes
C=C aromatic ring1600StrongAromatic quadrant stretch — universal aromatic marker
Amide I (protein)1655MediumC=O stretch — protein secondary structure
C≡N nitrile2200–2260StrongTriple-bond stretch — diagnostic silent-region peak
C-H aliphatic2850–2960StrongCH₂/CH₃ symmetric and asymmetric stretch
C-H aromatic3000–3100Mediumsp² C-H stretch — above 3000 cm⁻¹ distinguishes from sp³
How to use this table
  1. Find your peak position (Raman shift in cm⁻¹) from your spectrum.
  2. Type the wavenumber into the search box to highlight matching entries, or browse by material class.
  3. Use the intensity (strong, medium, weak) and context notes to narrow your assignment — a strong Raman peak from a symmetric vibration is more likely than a weak one.
  4. Confirm your assignment by checking for additional corroborating peaks from the same material listed in the description.

Organic Functional Groups

Aromatic Compounds

Polymers

Carbon Materials

Minerals & Inorganics

Biological Molecules

Solvents & Common Materials

Gases

Don't want to look up peaks manually? Upload your spectrum to the Raman Peak Identifier and get automated assignments with confidence scores.

Notes on Using This Table

  • Raman shift ranges are approximate — exact positions depend on the crystal structure, phase, strain state, temperature, and laser excitation wavelength (for resonance-enhanced modes).
  • Relative intensities are qualitative guides. Actual measured intensities depend on the laser wavelength, focusing conditions, sample orientation, and whether resonance enhancement occurs.
  • Some entries overlap in wavenumber range. Use additional peaks and context (sample type, other bands present) to distinguish between possible assignments.
  • Fluorescence background can obscure weak Raman bands. If you see a rising, curved baseline, consider baseline correction before peak identification.

Related Resources

Frequently Asked Questions

What is a Raman shift table?

A Raman shift table is a reference chart listing the characteristic Raman scattering frequencies (in cm⁻¹) for common functional groups, materials, and compounds. It helps spectroscopists identify unknown materials by matching observed peak positions to known assignments. Unlike IR correlation tables, Raman tables emphasize symmetric vibrations and polarizable bonds.

Why are some Raman peaks strong and others weak?

Raman peak intensity depends on the polarizability change during the vibration. Symmetric stretches of highly polarizable bonds (C=C, S-S, aromatic rings, C≡C) produce the strongest Raman bands. Asymmetric stretches and vibrations of polar bonds (O-H, N-H) are typically weak in Raman but strong in IR. This complementarity is why Raman and IR are often used together.

What is the silicon peak used for in Raman spectroscopy?

Crystalline silicon produces a sharp, intense Raman peak at 520.7 cm⁻¹ that is used as a universal calibration standard for Raman spectrometers. The peak position is precisely known and reproducible. In semiconductor research, shifts from 520.7 cm⁻¹ indicate mechanical stress or strain in the silicon crystal.